Sterilization & Disinfectant recommendations for
our products: |
|
The recommendations indicated below are to be used as
a general guideline. Follow the specific sterilization
or disinfection procedures that have been validated by
your institution.
Some
of Instrumentation Industries, Inc. products have sterilization
parameters listed on their instruction sheets. Follow
the recommendations on the instruction sheets for these
products.Always carefully examine
and test product after sterilization. Do not use product
if it shows signs of cracks, leaks, poor fit, severe discoloration,
or fails to perform as indicated. |
Reusable
Plastics & Silicone Rubber Products |
A
thorough pre-washing of parts is necessary to remove
foreign and/or organic contaminants. Use a low alkaline
soap (pH 8.5 or less) to preserve maximum useful life.
Pasteurization:
Pasteurize at 160F-170F (71C-77C) for no less than 30
minutes.
Ethylene
Oxide: Sterilize according to validated parameters.
Do not exceed 131F (55C). Allow ample aeration time
in a well ventilated area to dissipate the absorbed
gas.
Steam Autoclave: Sterilize according
to validated parameters. Do not exceed 275F (135C).
(Do not steam autoclave polycarbonate products)Gamma
Radiation: Irradiate according to validated parameters.
Some discoloration may occur. (Do not irradiate polyethylene
products).
Chemical Disinfectants: Disinfect according
to validated parameters. Follow the chemical manufacturer's
recommendation for temperature and soak time. Chemical
disinfection should be followed by a sterile water rinse.
Gas Plasma: Sterilize according to
validated parameters.
|
Disposable
Plastics |
Instrumentation
Industries, Inc. disposable products are designed for
single-patient use only. Pasteurization or chemical disinfection
is recommended.products.
Do not steam autoclave disposable products! |
Endotracheal
Tube Changers |
|
Endotracheal Tube Changers are single patient use only.
Sterilize in 100% ethylene oxide at 129°F (54°C)
for 60 minutes. Sterilized tubes must be allowed to
aerate for no less than 12 hours to dissipate any residual
gas.
Steam autoclave, gas plasma, sterrad, and gamma radiation
are NOT recommended processing methods.
|
KC
Reusable Tubing |
|
Tubing should be cleaned with a mild detergent followed
by a sterile water rinse prior to processing.
Reusable Tubing:
Steam Autoclave at 270F (132C)
Pasteurize at 170F (75C)
EtO at 131F (55C)
Gamma Radiation
Disposable Tubing:
EtO at 131F (55C)
Gamma Radiation
|
| Home
Use |
Wash parts in a mild liquid dish detergent. Parts should
be thoroughly scrubbed in order to remove all contaminants.
Rinse well; making sure all remaining detergent is removed.
Soak parts for 20 minutes in a fresh vinegar solution
that is 1 part vinegar and 3 parts water. (EXAMPLE: ½
cup vinegar and 1 ½ cup water) Thoroughly rinse
parts with tap water.
Air-dry on a clean towel. (Do not wipe or dry with towel) |
| Download a Printable version of iii
Sterilization & Disinfection Recommendations |
| |
|
 |
 |
 |
 |
 |
|
|
| |
Home
| About Us | All
Products | Search
| Customer Login
| View Cart
Dealers | Trade
Shows | OEM
| Sterilization Procedures |
Job Opportunities | Contact
Us | FAQ
©2010 Instrumentation Industries, Inc. All
Rights Reserved. - iiimedical.com
|
|
|

MDI Adapters. 
Leverhalers 
Gauge Stands
|