Sterilization & Disinfectant Recommendations |
|
The recommendations indicated below are to be used as
general guidelines. Follow the specific sterilization
or disinfection procedures that have been validated by
your institution.
Always carefully examine the physical integrity of the
product after processing. Do not use if cracking or crazing
is seen, or if parts fit together improperly after processing. |
Reusable
Plastics & Silicone Rubber Products |
| A
thorough pre-washing of parts is necessary to remove
foreign and/or organic contaminants.
Use a low alkaline soap (pH 8.5 or less) to preserve
maximum useful life.
| Sterilization: |
High
Level Disinfection: |
Steam
Autoclave:
Sterilize according to validated
parameters. Do not exceed 275°F (135°C).
Do not steam
autoclave polycarbonate products. |
Pasteurization:
Pasteurize at 153°F -
163°F (70°C +/ -3°) for minimum of 30
minutes. |
Gamma
Radiation :
Irradiate according to validated
parameters. Some discoloration may occur.
Do not irradiate
polyethylene products. |
Chemical
Disinfectants :
Disinfect according to
validated parameters.
Follow the chemical manufacturer's recommendation
for temperature and soak time.
Chemical disinfection should be followed by sterile
water rinse.
Recommended Chemical: 2-4% Activated Glutaraldehyde
Exposure time should be based on the
manufacturer's indication for use as a high-level
disinfectant or sterilant.
Do not use alcohol or chemicals containing dimethyl
ammonia chloride. |
| Ethylene
Oxide:
Sterilize according to
validated parameters. Do not exceed 131°F
(55°C).
Allow ample aeration time in a well-ventilated
area to dissipate the absorbed gas |
Gas
Plasma :
Sterilize according to validated
parameters. |
|
Disposable
-- Single Patient Use Products |
| Instrumentation
Industries, Inc. disposable products are designed for
single-patient use. Pasteurization or chemical disinfection
is recommended for most Single Patient Use products
prior to usage. Do not steam autoclave!
| Sterilization: |
High
Level Disinfection: |
STERRAD
100S: (Polystyrene Butadiene only)
Full short sterilization
cycle (Hydrogen gas plasma, 55 minutes, 45°C
- 55°C). |
Pasteurization:
Pasteurize at 153°F -
163°F (70°C +/ -3°) for a minimum of
30 minutes. |
| |
Cidex®
OPA
Prepare Cidex® OPA solution.
Use syringe to fill all crevices with Cidex®
OPA, then totally submerge adapters in the disinfectant.
Follow the manufacturer's instruction for temperature
and soak time requirements. |
|
Endotracheal
Tube Changers |
|
Endotracheal Tube Changers are designed for Single Patient
Use only. Recommended High Level Disinfection methods
are Pasteurization at 153°F - 163°F ) 70°C
+/ -3°C) for a minimum of 30 minutes, or Chemical
Disinfection. Sterilization can be performed in 100%
ethylene oxide at 129°F (54°C) for 60 minutes.
Sterilized tubes must be allowed to aerate for no less
12 hours to dissipate any residual gas.
Steam autoclave,
gas plasma and gas radiation are not recommended processing
methods.
|
Ventilator
Tubing |
Tubing
should be cleaned with a mild detergent followed by
a sterile water rinse prior to processing.
| Reusable
Tubing: |
Single
Patient Use Tubing: |
•
Steam Autoclave at 270°F (132°C)
• Pasteurize at 170°F (77°C)
• EtO at 131°F (55°C)
• Gamma Radiation |
•
EtO at 13°F (55C)
• Gamma Radiation |
|
| Metal
Products |
| Instrumentation
Industries, Inc. metal products include all brackets,
flex arms, instant flow valves, and manual jet ventilators.
Clean these products by wiping the outside with a disinfectant
on a soft cloth. Do not
immerse, disassemble, use bleach, or sterilize. |
| Home
Use |
Wash parts in a mild liquid dish detergent. Parts should
be thoroughly scrubbed in order to remove all contaminants.
Rinse well; ensure all remaining detergent is removed.
Soak parts for 20 minutes in a fresh vinegar solution
that is 1 part vinegar and 3 parts water. (EXAMPLE: ½
cup vinegar and 1 ½ cup water)
Thoroughly rinse
parts with water. Allow to air-dry on a clean towel. Do
not wipe or dry with towel. |
| Download a Printable version of iii
Sterilization & Disinfection Recommendations |
| |
|
 |
 |
 |
 |
 |
|
|
| |
Home
| About Us | All
Products | Search
| Customer Login
| View Cart
Dealers | Trade
Shows | OEM
| Sterilization Procedures |
Job Opportunities | Contact
Us | FAQ
©2012
Instrumentation Industries, Inc. All Rights Reserved. - iiimedical.com
|
|
|

Universal Inline
MDI Adapters

Pilot Tube Repair Kits 
Gauge Stands
|