Sterilization & Disinfectant recommendations for our products:
The recommendations indicated below are to be used as a general guideline. Follow the specific sterilization or disinfection procedures that have been validated by your institution.   Some of Instrumentation Industries, Inc. products have sterilization parameters listed on their instruction sheets. Follow the recommendations on the instruction sheets for these products.Always carefully examine and test product after sterilization. Do not use product if it shows signs of cracks, leaks, poor fit, severe discoloration, or fails to perform as indicated.
Reusable Plastics & Silicone Rubber Products
A thorough pre-washing of parts is necessary to remove foreign and/or organic contaminants. Use a low alkaline soap (pH 8.5 or less) to preserve maximum useful life. Pasteurization: Pasteurize at 160F-170F (71C-77C) for no less than 30 minutes.Steam Autoclave: Sterilize according to validated parameters. Do not exceed 275F (135C). (Do not steam autoclave polycarbonate products)Gamma Radiation: Irradiate according to validated parameters. Some discoloration may occur. (Do not irradiate polyethylene products)Ethylene Oxide: Sterilize according to validated parameters. Do not exceed 131F (55C). Allow ample aeration time in a well ventilated area to dissipate the absorbed gas.Chemical Disinfectants: Disinfect according to validated parameters. Follow the chemical manufacturer's recommendation for temperature and soak time. Chemical disinfection should be followed by a sterile water rinse.Gas Plasma: Sterilize according to validated parameters.
Disposable Plastics
Instrumentation Industries, Inc. disposable products are designed for single-patient use only. Pasteurization or chemical disinfection is recommended.products.
Do not steam autoclave disposable products!
Endotracheal Tube Changers

Endotracheal Tube Changers are single-patient use only.
Pasteurization or chemical
disinfection is recommended.
Steam autoclave, gas plasma and gamma radiation are
NOT recommended processing methods.

KC Reusable Tubing
Tubing should be cleaned with a mild detergent followed by a sterile water rinse prior to processing.
Reusable Tubing:
Steam Autoclave at 270F (132C)
Pasteurize at 170F (75C)
EtO at 131F (55C)
Gamma Radiation

Disposable Tubing:
EtO at 131F (55C)
Gamma Radiation
Home Use
Wash parts in a mild liquid dish detergent. Parts should be thoroughly scrubbed in order to remove all contaminants. Rinse well; making sure all remaining detergent is removed. Soak parts for 20 minutes in a fresh vinegar solution that is 1 part vinegar and 3 parts water. (EXAMPLE: ½ cup vinegar and 1 ½ cup water) Thoroughly rinse parts with tap water.
Air-dry on a clean towel. (Do not wipe or dry with towel)
Download a Printable version of iii Sterilization & Disinfection Recommendations
 

 
 