Login Status |
|
Not logged in
|
| » Login |
Order
Direct from Us or from one of our worldwide
DEALERS
We accept:

We also accept: Purchase Orders
from Government, Hospitals, Schools, Universities, Fortune 1000 Companies,
and International orders.
Download our
PRICELIST (.pdf)
To join our mailing list click HERE
Download
this month's featured products
Brochure
PDF

Instrumentation
Industries, Inc.
2990 Industrial Blvd.
Bethel Park. PA 15102
Friendly
Customer Service:
(M-F,
8am-5pm, EST)
1-800-633-8577
1-412-854-1133
Fax Number:
1-412-854-5668
EMAIL


ISO
13485: 2003 Certified. FM 72200

|
|
| |
 |
 |
 |
|
Sterilization & Disinfectant recommendations for
our products: |
|
The recommendations indicated below are to be used as
a general guideline. Follow the specific sterilization
or disinfection procedures that have been validated by
your institution.
Some
of Instrumentation Industries, Inc. products have sterilization
parameters listed on their instruction sheets. Follow
the recommendations on the instruction sheets for these
products.Always carefully examine
and test product after sterilization. Do not use product
if it shows signs of cracks, leaks, poor fit, severe discoloration,
or fails to perform as indicated. |
Reusable
Plastics & Silicone Rubber Products |
| A
thorough pre-washing of parts is necessary to remove foreign
and/or organic contaminants. Use a low alkaline soap (pH
8.5 or less) to preserve maximum useful life.
Pasteurization: Pasteurize at 160F-170F
(71C-77C) for no less than 30 minutes.Steam
Autoclave: Sterilize according to validated parameters.
Do not exceed 275F (135C). (Do not steam autoclave polycarbonate
products)Gamma Radiation: Irradiate
according to validated parameters. Some discoloration
may occur. (Do not irradiate polyethylene products)Ethylene
Oxide: Sterilize according to validated parameters. Do
not exceed 131F (55C). Allow ample aeration time in a
well ventilated area to dissipate the absorbed gas.Chemical
Disinfectants: Disinfect according to validated parameters.
Follow the chemical manufacturer's recommendation for
temperature and soak time. Chemical disinfection should
be followed by a sterile water rinse.Gas
Plasma: Sterilize according to validated parameters. |
Disposable
Plastics |
Instrumentation
Industries, Inc. disposable products are designed for
single-patient use only. Pasteurization or chemical disinfection
is recommended.products.
Do not steam autoclave disposable products! |
Endotracheal
Tube Changers |
|
Endotracheal Tube Changers are single-patient use only.
Pasteurization or chemical
disinfection is recommended.
Steam autoclave, gas plasma and gamma radiation are
NOT recommended processing methods.
|
KC
Reusable Tubing |
|
Tubing should be cleaned with a mild detergent followed
by a sterile water rinse prior to processing.
Reusable Tubing:
Steam Autoclave at 270F (132C)
Pasteurize at 170F (75C)
EtO at 131F (55C)
Gamma Radiation
Disposable Tubing:
EtO at 131F (55C)
Gamma Radiation
|
| Home
Use |
Wash parts in a mild liquid dish detergent. Parts should
be thoroughly scrubbed in order to remove all contaminants.
Rinse well; making sure all remaining detergent is removed.
Soak parts for 20 minutes in a fresh vinegar solution
that is 1 part vinegar and 3 parts water. (EXAMPLE: ½
cup vinegar and 1 ½ cup water) Thoroughly rinse
parts with tap water.
Air-dry on a clean towel. (Do not wipe or dry with towel) |
| Download a Printable version of iii
Sterilization & Disinfection Recommendations |
| |
|
 |
 |
 |
 |
 |
|
|
| |
Home
| About Us | All
Products | Search
| Customer Login
| View Cart
Dealers | Trade
Shows | OEM
| Sterilization Procedures |
Job Opportunities | Contact
Us
©2007 Instrumentation Industries, Inc. All
Rights Reserved. - iiimedical.com
|
|
|

MDI Adapters. 
Leverhalers 
Gauge Stands
|